Title of article
Dosimetric properties of a radiochromic gel detector for diagnostic X-rays
Author/Authors
Bero، نويسنده , , M.A.، نويسنده ,
Pages
4
From page
186
To page
189
Abstract
The gel dosimetry method was found to be capable of addressing complicated issues related to dose measurements particularly in modern sophisticated radiotherapy applications. The Ferrous-sulphate Xylenol-orange and Gelatin (FXG) radiochromic gel dosemeter is one of the systems used for such applications. Some chemical dosemeters show different responses for low- and medium-energy X-rays in comparison with high-energy γ-photons. The energy and dose rate dependence of the FXG dose response was examined. In addition to the detector response, other important dosimetric properties of the system were investigated for different X-ray beam qualities with tube voltages in the range 100–300 kV. An orthovoltage X-ray therapy unit was used to irradiate standard sized samples of FXG from different batches for radiation doses in the range 0–20 Gy.
Keywords
Radiochromic gel detector , X-ray beam qualities , Radiation dosimetry
Journal title
Astroparticle Physics
Record number
2029965
Link To Document