Title of article
Measurement and Monte Carlo modeling of the spatial response of scintillation screens
Author/Authors
Pistrui-Maximean، نويسنده , , S.A. and Létang، نويسنده , , J.M. and Freud، نويسنده , , N. and Koch، نويسنده , , A. and Walenta، نويسنده , , A.H. and Montarou، نويسنده , , G. and Babot، نويسنده , , D.، نويسنده ,
Pages
9
From page
719
To page
727
Abstract
In this article, we propose a detailed protocol to carry out measurements of the spatial response of scintillation screens and to assess the agreement with simulated results. The experimental measurements have been carried out using a practical implementation of the slit method. A Monte Carlo simulation model of scintillator screens, implemented with the toolkit Geant4, has been used to study the influence of the acquisition setup parameters and to compare with the experimental results. An algorithm of global stochastic optimization based on a localized random search method has been implemented to adjust the optical parameters (optical scattering and absorption coefficients). The algorithm has been tested for different X-ray tube voltages (40, 70 and 100 kV). A satisfactory convergence between the results simulated with the optimized model and the experimental measurements is obtained.
Keywords
x-ray imaging , Monte Carlo simulation , GEANT4 , Slit method , modulation transfer function (MTF) , Phosphor screen
Journal title
Astroparticle Physics
Record number
2030969
Link To Document