Comparison of experimental and theoretical efficiency of HPGe X-ray detector
Author/Authors
Mohanty، نويسنده , , B.P. and Balouria، نويسنده , , Ravi P. and Garg، نويسنده , , M.L. and Nandi، نويسنده , , T.K. and Mittal، نويسنده , , V.K. and Govil، نويسنده , , I.M.، نويسنده ,
Pages
5
From page
186
To page
190
Abstract
The low energy high purity germanium (HPGe) detectors are being increasingly used for the quantitative estimation of elements using X-ray spectrometric techniques. The softwares used for quantitative estimation normally evaluate model based efficiency of detector using manufacturer supplied detector physical parameters. The present work shows that the manufacturer supplied detector parameters for low energy HPGe detectors need to be verified by comparing model based efficiency with the experimental ones. This is particularly crucial for detectors with ion implanted P type contacts.
Keywords
HPGe detector , X-ray spectrometry , efficiency , Si(Li) detector
Journal title
Astroparticle Physics
Record number
2031395