• Title of article

    Measurement of the thermal conductivity of argon and methane: a test of a transient hot-wire apparatus

  • Author/Authors

    Pلtek، نويسنده , , J and Klomfar، نويسنده , , J، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2002
  • Pages
    17
  • From page
    147
  • To page
    163
  • Abstract
    A transient hot-wire apparatus for absolute measurements of the thermal conductivity of gases has been constructed and tested with argon and methane. The apparatus operates at temperatures from about 300 to 473 K and pressures up to 60 MPa. To evaluate the capability of the apparatus, measurements were performed (i) with argon along five isotherms at temperatures between 300 and 425 K with pressures up to 15 MPa, and (ii) with methane along seven isotherms at temperatures between 290 and 360 K with pressures up to 15 MPa. The experimental thermal conductivity data and the results of their low-density analysis are compared with results of other investigators. The total experimental uncertainty of the measurements is estimated to be ±0.001×p in pressure and ±0.005 K in temperature. The relative uncertainty of the measurements at 95% level is estimated to be 1.2% for argon and 0.7% for methane. The reproducibility of the results measured by the standard deviation from the polynomial λ(ρ) relation along isotherms is 0.3% for argon and 0.2% for methane.
  • Keywords
    thermal conductivity , Transient hot-wire apparatus , Pure , Methane , argon
  • Journal title
    Fluid Phase Equilibria
  • Serial Year
    2002
  • Journal title
    Fluid Phase Equilibria
  • Record number