• Title of article

    On specimen tilt for electron holography of semiconductor devices

  • Author/Authors

    Formanek، نويسنده , , Petr and Bugiel، نويسنده , , Eberhard، نويسنده ,

  • Pages
    9
  • From page
    292
  • To page
    300
  • Abstract
    Electron holography can be successfully used for potential mapping on a nanometer scale. It relies on the fact that the phase of the electron wave is proportional to the electrostatic potential in the specimen. However, this proportionality is valid only in a kinematical condition, which is achieved by proper specimen orientation with respect to the electron beam. In this report, we examine experimentally in detail the specimen orientations of silicon devices that minimize dynamical diffraction. The tilt of the specimen from the edge-on position to a favorable orientation causes certain interfaces in the specimen to smear. We describe the smearing by a transfer function and compare it to wave transfer function of the microscope. The maximal specimen tilt αmax that does not cause smearing greater than the resolution r of the microscope is αmax=arctan(5.70r/t), where t is the specimen thickness.
  • Keywords
    Electron holography , Dopant profiling , Kinematical condition , Dynamical condition , Specimen tilt
  • Journal title
    Astroparticle Physics
  • Record number

    2051843