• Title of article

    Room design for high-performance electron microscopy

  • Author/Authors

    Muller، نويسنده , , David A. and Kirkland، نويسنده , , Earl J. and Thomas، نويسنده , , Malcolm G. and Grazul، نويسنده , , John L. and Fitting، نويسنده , , Lena and Weyland، نويسنده , , Matthew، نويسنده ,

  • Pages
    8
  • From page
    1033
  • To page
    1040
  • Abstract
    Aberration correctors correct aberrations, not instabilities. Rather, as spatial resolution improves, a microscopeʹs sensitivity to room environment becomes more noticeable, not less. Room design is now an essential part of the microscope installation process. Previously ignorable annoyances like computer fans, desk lamps and that chiller in the service corridor now may become the limiting factors in the microscopes performance. We discuss methods to quantitatively characterize the instrumentʹs response to magnetic, mechanical, acoustical and thermal disturbances and thus predict the limits that the environment places on imaging and spectroscopy.
  • Keywords
    High-performance electron microscopy
  • Journal title
    Astroparticle Physics
  • Record number

    2051929