Title of article
Wear performance of laser processed tantalum coatings
Author/Authors
Dittrick، نويسنده , , Stanley and Balla، نويسنده , , Vamsi Krishna and Bose، نويسنده , , Susmita and Bandyopadhyay، نويسنده , , Amit، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2011
Pages
4
From page
1832
To page
1835
Abstract
This first generation investigation evaluates the in vitro tribological performance of laser-processed Ta coatings on Ti for load-bearing implant applications. Linear reciprocating wear tests in simulated body fluid showed one order of magnitude less wear rate, of the order of 10− 4 mm3(N.m)− 1, for Ta coatings compared to Ti. Our results demonstrate that Ta coatings can potentially minimize the early-stage bone-implant interface micro-motion induced wear debris generation due to their excellent bioactivity comparable to that of hydroxyapatite (HA), high wear resistance and toughness compared to popular HA coatings.
Keywords
IN VITRO , Laser deposition , WEAR , biomaterial , tantalum