• Title of article

    Corrosion mechanism applicable to biodegradable magnesium implants

  • Author/Authors

    Atrens، نويسنده , , Andrej and Liu، نويسنده , , Ming and Zainal Abidin، نويسنده , , Nor Ishida، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2011
  • Pages
    28
  • From page
    1609
  • To page
    1636
  • Abstract
    Much of our understanding of the Mg corrosion mechanism is based on research using aggressive chloride based solutions like 3% NaCl, which are appropriate for understand the corrosion for applications such as auto construction. The chloride ions tend to cause break down of the partly protective surface film on the Mg alloy surface. The corrosion rate increases with exposure time until steady state is reached, which may take several weeks. An overview is provided of the aspects which determine the corrosion of Mg alloys: (i) measurement details; (ii) impurity elements Fe, Ni, Cu and Co; (iii) second phases; (iv) surface films and surface condition and (v) stress corrosion cracking (SCC). This understanding is used to help understand Mg corrosion for Mg as a biodegradable implant for medical applications. Solutions that elucidate these applications tend to form surface films and the corrosion rate tends to decrease with immersion time.
  • Keywords
    Corrosion , Stress corrosion cracking , Biodegradable implants , Magnesium
  • Journal title
    MATERIALS SCIENCE & ENGINEERING: B
  • Serial Year
    2011
  • Journal title
    MATERIALS SCIENCE & ENGINEERING: B
  • Record number