• Title of article

    A simple high-sensitivity technique for purity analysis of xenon gas

  • Author/Authors

    Leonard، نويسنده , , D.S. and Dobi، نويسنده , , A. and Hall، نويسنده , , L.J. Kaufman، نويسنده , , C. and Langford، نويسنده , , T. and Slutsky، نويسنده , , S. and Yen، نويسنده , , Y.-R.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2010
  • Pages
    7
  • From page
    678
  • To page
    684
  • Abstract
    We report on the development and performance of a high-sensitivity purity-analysis technique for gaseous xenon. The gas is sampled at macroscopic pressure from the system of interest using a UHV leak valve. The xenon present in the sample is removed with a liquid-nitrogen cold trap, and the remaining impurities are observed with a standard vacuum mass-spectroscopy device. Using calibrated samples of xenon gas spiked with known levels of impurities, we find that the minimum detectable levels of N2, O2, and methane are 1×10−9, 160×10−12, and 60 ×10−12 g/g, respectively. This represents an improvement of about a factor of 10 000 compared to measurements performed without a cold trap.
  • Keywords
    Purification , Cold trap , mass spectrometry , Mass Spectroscopy , Noble gas , Xenon
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Serial Year
    2010
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Record number

    2170474