Title of article
Sensitivity determination and optimisation of a cylindrical diffusion chamber, for radon measurements, with a CR39 detector
Author/Authors
Nikezi?، نويسنده , , Dragoslav and Baixeras، نويسنده , , Carmen and Kosti?، نويسنده , , Dragana، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 1996
Pages
9
From page
290
To page
298
Abstract
This paper deals with the determination of the sensitivity of the CR39 detector for radon measurements. The sensitivity was calculated by applying the Monte Carlo method for a detector in a diffusion cylindrical chamber with different radii and heights. The sensitivity is reduced as the chamber dimension decreased.
certainty of radon concentration determination associated with the unknown radon progeny deposition on the inner chamber walls is discussed. This uncertainty can be almost completely avoided by using a chamber narrower than 1.25 cm and higher than 5 cm. Such design of the diffusion chamber can enable better and more accurate radon measurements to be achieved.
Journal title
Nuclear Instruments and Methods in Physics Research Section A
Serial Year
1996
Journal title