Title of article
Comparative evaluation of the temperature dependence of different noise sources in CdTe detectors
Author/Authors
Chirco، نويسنده , , P and Zanarini، نويسنده , , M and Caroli، نويسنده , , E and Donati، نويسنده , , A and Dusi، نويسنده , , W and Stephen، نويسنده , , J.B and DiCocco، نويسنده , , G، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 1996
Pages
5
From page
127
To page
131
Abstract
The number of applications of CdTe detectors for imaging purposes is currently growing at a fast rate. Nevertheless, noise is still a limiting factor in several circumstances, particularly when a high dynamic range is needed. Thus, knowledge of the comparative importance of various noise sources (mainly referring to the dark current in the detector, the electronic noise in the preamplifier and the statistical noise) is valuable for the development of efficient noise reduction techniques. The dependence on temperature of these noise sources has been evaluated over the range from −30°C up to 40°C. Although some components of the noise show a marked temperature dependence, the overall detector spectroscopic performances are constant up to 10°C, while above this level an appreciable degradation occurs.
Journal title
Nuclear Instruments and Methods in Physics Research Section A
Serial Year
1996
Journal title
Nuclear Instruments and Methods in Physics Research Section A
Record number
2173677