    X-ray microtomography scanner using time-delay integration for elimination of ring artefacts in the reconstructed image

  • Author/Authors

    Davis، نويسنده , , G.R and Elliott، نويسنده , , J.C، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 1997
  • Pages
    6
  • From page
    157
  • To page
    162
  • Abstract
    Most X-ray microtomography scanners work on the same principle as third-generation medical CT scanners, that is, the same point in each projection is measured by the same detector element. This leads to ring artefacts in the reconstructed image if the X-ray sensitivities of the individual detector elements, after any analytical correction, are not all identical. We have developed an X-ray microtomography scanner which uses the time-delay integration method of imaging with a CCD detector to average the characteristics of all the detector elements in each linear projection together. This has the added advantage of allowing specimens which are larger than the detector and X-ray field to be scanned. The device also uses a novel mechanical stage to “average out” inhomogeneities in the X-ray field. The results show that ring artefacts in microtomographic images are eliminated using this technique.
  • Keywords
    Microtomography , Ring artefacts , CCD camera , Time-delay integration