• Title of article

    Electrode thickness measurement of a Si(Li) detector for the SIXA array

  • Author/Authors

    Tikkanen، نويسنده , , T. and Hنmنlنinen، نويسنده , , K. Alanko-Huotari، نويسنده , , S.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 1998
  • Pages
    6
  • From page
    425
  • To page
    430
  • Abstract
    Cathode electrodes of the Si(Li) detector elements of the SIXA X-ray spectrometer array are formed by gold-palladium alloy contact layers. The equivalent thickness of gold in one element was measured by observing the characteristic L-shell X-rays of gold excited by monochromatised synchrotron radiation with photon energies above the L3 absorption edge of gold. The results obtained at four different photon energies below the L2 edge yield an average value of 22.4 ± 3.5 nm which is consistent with the earlier result extracted from detection efficiency measurements.
  • Keywords
    Gold electrodes , Synchrotron radiation , Si(Li) detectors , X-ray spectrometers , X-ray fluorescence , detector calibration
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Serial Year
    1998
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Record number