• Title of article

    Use of photodiode in microdosimetry and evaluation of effective quality factor

  • Author/Authors

    Kadachi، نويسنده , , Syed A. and Waheed، نويسنده , , A. and Al-Eshaikh، نويسنده , , M. and Obeid، نويسنده , , M.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 1998
  • Pages
    7
  • From page
    400
  • To page
    406
  • Abstract
    A silicon photodiode in conjunction with a charged particle radiator placed in front of the sensitive layer was employed to construct a microdosimetric detector. Silicon photodiodes of different types (PN, PIN and PNN) and of different sensitive layers (1, 3, 5 and 8 μm) were used to obtain the pulse height spectrum for an AmBe neutron source. The data were analysed to determine the microdosimetric spectra and the effective quality factor Q was estimated. Two types of corrections were needed for the estimate of Q. One is related to the weighting function Q(L) and the other deals with the fractional relative dose d(L). The results were compared with those obtained with Tissue Equivalent Proportional Counter (TEPC) of 1 μm simulated diameter and were found to be in good agreement.
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Serial Year
    1998
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Record number

    2177318