• Title of article

    Electrical measurement of wire tension in a multi-wire drift chamber

  • Author/Authors

    Ohama، نويسنده , , John T. and Ishihara، نويسنده , , N. and Takeda، نويسنده , , S. and Okuma، نويسنده , , H. and Konno، نويسنده , , K.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 1998
  • Pages
    4
  • From page
    175
  • To page
    178
  • Abstract
    A new and simple technique has been developed to measure the wire tension of a multi-wire drift chamber, which is the prototype chamber of DCBA (Drift Chamber Beta-ray Analyzer) for double beta decay experiments. The present method uses two wires adjacent to each other in the chamber for applying AC and DC electric fields, and can easily detect a weak current signal at the resonance frequency corresponding to the wire tension. The wire tension has been measured with an accuracy of about 1%.
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Serial Year
    1998
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Record number

    2178209