Title of article
The study of α/γ ratio for inorganic scintillation detectors
Author/Authors
Sysoeva، نويسنده , , E.V. and Tarasov، نويسنده , , V.A and Zelenskaya، نويسنده , , O.V and Sulyga، نويسنده , , V.A، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 1998
Pages
5
From page
274
To page
278
Abstract
To optimize characteristics of α-radiation detectors as well as those intended to stabilize γ spectrometers, the α/γ ratio has been studied for inorganic scintillation detectors within a wide temperature range (from −20 to +80°C) and at various signal formation time values (from 0.2 to 20 μs). Inorganic scintillators with high atomic numbers have been studied, namely, alkali halide and oxide crystals as well as a semiconducting single-crystal ZnSe(Te). Though the scintillators on the base of the oxide materials have satisfactory resolution and light output, they cannot be considered as efficient detectors of α-radiation due to a low α/γ ratio (∼0.20). The most appropriate detector of α-radiation remains single crystalline CsI(Tl) having the best scintillation characteristics, nonhygroscopic, convenient to use and comparatively cheap.
Keywords
Signal formation time , Energy resolution , temperature dependence , scintillation detector , Light output , ?/? Ratio
Journal title
Nuclear Instruments and Methods in Physics Research Section A
Serial Year
1998
Journal title
Nuclear Instruments and Methods in Physics Research Section A
Record number
2179084
Link To Document