Title of article
The charged particle response of CdZnTe radiation detectors
Author/Authors
Morton، نويسنده , , E.J. and Crockett، نويسنده , , G.M. and Sellin، نويسنده , , P.J. and DeAntonis، نويسنده , , P.، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 1999
Pages
4
From page
169
To page
172
Abstract
A series of experiments has determined the response of CdZnTe radiation detectors to electrons and alpha particles. Densely ionising particles result in an “S-shaped” pulse which is somewhat different in shape to those generated during photon interactions. A preliminary analysis of these pulse shapes is presented and the impact of this on possible schemes for pulse-shape discrimination is discussed. By altering the position of interaction of the alpha-particle beam in the detector, it is possible to analyse the non-uniformity of the electric field present in these detectors. Results are presented to demonstrate these effects, and the implications of these for operation of CdZnTe detectors in charged particle beams will be discussed.
Journal title
Nuclear Instruments and Methods in Physics Research Section A
Serial Year
1999
Journal title
Nuclear Instruments and Methods in Physics Research Section A
Record number
2180377
Link To Document