• Title of article

    Portable X-ray and gamma-ray imager with coded mask: performance characteristics and methods of image reconstruction

  • Author/Authors

    Ivanov، نويسنده , , O.P and Sudarkin، نويسنده , , A.N and Stepanov، نويسنده , , V.E. and Urutskoev، نويسنده , , L.I، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 1999
  • Pages
    6
  • From page
    729
  • To page
    734
  • Abstract
    The system is based on combined detector: scintillating plate+image intensifier+CCD matrix. The sensitivity of the detector allows the detection of single photon with energy higher than 60 keV. The hexagonal URA masks of ranks 6 and 9 are used for coded mask imaging; transparent elements of mask are cylindrical holes of diameter 1.3–2.2 mm). The imaging possibilities of device are presented by images of point and circular objects of different diameters obtained in experiments.
  • Keywords
    X-ray and gamma-ray imager , Image reconstruction , HERV
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Serial Year
    1999
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Record number

    2180483