• Title of article

    Absolute calibration of photomultiplier based detectors – difficulties and uncertainties

  • Author/Authors

    Wright، نويسنده , , A.G.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 1999
  • Pages
    6
  • From page
    507
  • To page
    512
  • Abstract
    Photomultiplier manufacturers can provide a calibration of quantum efficiency over a range of wavelengths with an accuracy of up to 2%. To convert these figures to absolute counting efficiency requires knowledge of photomultiplier collection efficiency, F. Traditional methods for determining F are discussed with emphasis on sources of error. Light sources emitting at a known photon rate allow the absolute quantum efficiency to be determined directly. It is important in all attempts at absolute calibration to appreciate the conditions which manufacturers apply when calibrating photomultipliers.
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Serial Year
    1999
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Record number

    2182430