Title of article
X-ray CCD response functions, front to back
Author/Authors
Bautz، نويسنده , , M.W and Prigozhin، نويسنده , , G.Y and Pivovaroff، نويسنده , , M.J. and Jones، نويسنده , , S.E and Kissel، نويسنده , , S.E and Ricker، نويسنده , , G.R، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 1999
Pages
13
From page
40
To page
52
Abstract
For some applications, particularly in X-ray astronomy, one requires accurate knowledge of the charge-coupled device (CCD) spectral redistribution function, with errors of order 1%. We describe some recent advances in response function measurement and modeling. In particular, we show that it is now possible to isolate and separately measure the response contributions of many of the component volumes that comprise the CCD. We describe how we have deployed this capability to determine X-ray photon response functions for front-illuminated detectors, and indicate how these measurements can improve understanding of the response function of the back-illuminated CCD.
Journal title
Nuclear Instruments and Methods in Physics Research Section A
Serial Year
1999
Journal title
Nuclear Instruments and Methods in Physics Research Section A
Record number
2183414
Link To Document