Abstract
The installation for carrying out the investigations of synchrotron radiation (SR) and of electron beam diagnostics, Mossbauer and XAFS experiments was made. The installation is multipurpose due to its block construction. It consists of two precision double-crystal monochromators (range of rotation — 360°, precision — 5″, step — 0.545″), X-ray goniometer (range of rotation — 360°, precision — 18″, step — 3″), control and registration systems (time resolution 5 ns). It is possible to have fixed-in-space monochromatic SR beam.
Keywords
Mossbauer , XAFS , SR , Electron beam diagnostics