    X-ray CCD calibration system using fluorescent lines

  • Author/Authors

    Hamaguchi، نويسنده , , K. and Maeda، نويسنده , , Y. and Matsumoto، نويسنده , , H. and Nishiuchi، نويسنده , , M. and Tomida، نويسنده , , H. and Koyama، نويسنده , , K. and Awaki، نويسنده , , H. and Tsuru، نويسنده , , T.G.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2000
  • Pages
    5
  • From page
    360
  • To page
    364
  • Abstract
    We have developed a CCD calibration system using fluorescent X-ray lines with energies ranging from 1.49 keV (Al Kα) to 11.2 keV (Se Kα). The absolute X-ray flux is calibrated by a gas proportional counter, while the emerging spectra are monitored by solid-state silicon detectors. In order to suppress contaminating X-rays in the fluorescence spectra, mechanical collimators were set in the X-ray beam line, high-purity targets for fluorescent lines were used, and band-pass filters were put on the X-ray beam line. As for the purity of the fluorescent X-rays, the typical purity achieved was ∼98%.
  • Keywords
    CCD , Fluorescent X-rays , ASTRO-E , Calibration system