Title of article
Absolute energy calibration for plane grating monochromators
Author/Authors
Weiss، نويسنده , , M.R. and Follath، نويسنده , , R. and Sawhney، نويسنده , , K.J.S. and Zeschke، نويسنده , , T.، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2001
Pages
3
From page
482
To page
484
Abstract
This report describes a new method for monochromator energy calibration, which is based on the freedom to change the fix focus constant cff in the plane grating monochromator at BESSY. If the monochromator is not perfectly aligned, spectra taken at different cff-values show energy shifts (of 1–2%). These can be attributed to offsets in the grating angles. Measuring spectra at different cff-values, one can determine these offsets by employing a fitting procedure. If these offsets are inserted, cff variation measurements lead to relative energy scale changes of less than a few 10−4.
Keywords
Plane grating monochromators , Soft X-ray monochromator
Journal title
Nuclear Instruments and Methods in Physics Research Section A
Serial Year
2001
Journal title
Nuclear Instruments and Methods in Physics Research Section A
Record number
2191909