• Title of article

    Advance in a nano-accuracy surface profiler with an extended-angle test range

  • Author/Authors

    Qian، نويسنده , , Shinan and Qian، نويسنده , , Kun and Idir، نويسنده , , Mourad، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2013
  • Pages
    7
  • From page
    52
  • To page
    58
  • Abstract
    An advanced design of a nano-accuracy surface profiler (NSP) is introduced wherein we combined a scanning optical head with non-tilted reference system to facilitate measurements over an extended range of angles. The lateral motion of the beam during testing of a strongly curved mirror induces a systematic error. For a pencil-beam scanning profiler, the arm with varying optical-path lengths should be non-tilted so to eliminate the beamʹs lateral motion, and the arm for testing larger angles should be short and fixed so to reduce the beamʹs lateral motion. Our new scheme of having a non-tilted reference system offers an effective, simple, and convenient solution. A beam spot of 0.5–1 mm is used for higher spatial frequency tests in surface-figure measurements. Some preliminary studies and test are demonstrated.
  • Keywords
    Metrology , Nano-accuracy , Strongly curved mirror , Beam lateral motion , Profiler
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Serial Year
    2013
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Record number