• Title of article

    Determining 252Cf source strength by absolute passive neutron correlation counting

  • Author/Authors

    Croft، نويسنده , , S. and Henzlova، نويسنده , , D.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2013
  • Pages
    8
  • From page
    5
  • To page
    12
  • Abstract
    Physically small, lightly encapsulated, radionuclide sources containing 252Cf are widely used for a vast variety of industrial, medical, educational and research applications requiring a convenient source of neutrons. For many quantitative applications, such as detector efficiency calibrations, the absolute strength of the neutron emission is needed. In this work we show how, by using a neutron multiplicity counter the neutron emission rate can be obtained with high accuracy. This provides an independent and alternative way to create reference sources in-house for laboratories such as ours engaged in international safeguards metrology. The method makes use of the unique and well known properties of the 252Cf spontaneous fission system and applies advanced neutron correlation counting methods. We lay out the foundation of the method and demonstrate it experimentally. We show that accuracy comparable to the best methods currently used by national bodies to certify neutron source strengths is possible.
  • Keywords
    Multiplicity counting , Absolute metrology , Californium neutrons , Neutron correlation analysis , Source calibration
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Serial Year
    2013
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Record number