• Title of article

    Development of Multi-gap Resistive Plate Chamber (MRPC) for medical imaging

  • Author/Authors

    Banerjee، نويسنده , , A. and Roy، نويسنده , , A. and Biswas، نويسنده , , S. and Chattopadhyay، نويسنده , , S. and Das، نويسنده , , G. and Pal، نويسنده , , S.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2013
  • Pages
    2
  • From page
    138
  • To page
    139
  • Abstract
    The low cost and high resolution Multi-gap Resistive Plate Chamber (MRPC) opens up a new possibility to find an efficient alternative detector for the Time of Flight (TOF) based Positron Emission Tomography, where the sensitivity of the system depends largely on the time resolution of the detector. In a layered structure, suitable converters can be used to increase the photon detection efficiency. In this paper results of the cosmic ray test of a four-gap bakelite-based prototype MRPC operated in streamer mode and six-gap glass-based MRPC operated in avalanche mode are discussed.
  • Keywords
    Avalanche mode , MRPC , silicone , Cosmic rays , Time resolution
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Serial Year
    2013
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Record number

    2194156