Title of article
Development of Multi-gap Resistive Plate Chamber (MRPC) for medical imaging
Author/Authors
Banerjee، نويسنده , , A. and Roy، نويسنده , , A. and Biswas، نويسنده , , S. and Chattopadhyay، نويسنده , , S. and Das، نويسنده , , G. and Pal، نويسنده , , S.، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2013
Pages
2
From page
138
To page
139
Abstract
The low cost and high resolution Multi-gap Resistive Plate Chamber (MRPC) opens up a new possibility to find an efficient alternative detector for the Time of Flight (TOF) based Positron Emission Tomography, where the sensitivity of the system depends largely on the time resolution of the detector. In a layered structure, suitable converters can be used to increase the photon detection efficiency. In this paper results of the cosmic ray test of a four-gap bakelite-based prototype MRPC operated in streamer mode and six-gap glass-based MRPC operated in avalanche mode are discussed.
Keywords
Avalanche mode , MRPC , silicone , Cosmic rays , Time resolution
Journal title
Nuclear Instruments and Methods in Physics Research Section A
Serial Year
2013
Journal title
Nuclear Instruments and Methods in Physics Research Section A