• Title of article

    Feynman-α technique for measurement of detector dead time using a 30 kW tank-in-pool research reactor

  • Author/Authors

    Akaho، نويسنده , , E.H.K. and Intsiful، نويسنده , , J.D.K. and Maakuu، نويسنده , , B.T. and Anim-Sampong، نويسنده , , S. and Nyarko، نويسنده , , B.J.B.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2002
  • Pages
    6
  • From page
    585
  • To page
    590
  • Abstract
    Reactor noise analysis was carried out for Ghana Research Reactor-1 GHARR-1, a tank-in-pool type reactor using the Feynman-α technique (variance-to-mean method). Measurements made at different detector positions and under subcritical conditions showed that the technique could not be used to determine the prompt decay constant for the reactor which is Be reflected with photo-neutron background. However, for very low dwell times the technique was used to measure the dead time of the detector which compares favourably with the value obtained using the α-conventional method.
  • Keywords
    Feynman-? technique , Reactor noise analysis , Detector dead time , Dwell time
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Serial Year
    2002
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Record number

    2194744