• Title of article

    Study of a Cherenkov TOF-PET module

  • Author/Authors

    Korpar، نويسنده , , S. and Dolenec، نويسنده , , R. and Kri?an، نويسنده , , P. and Pestotnik، نويسنده , , R. and Stanovnik، نويسنده , , A.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2013
  • Pages
    4
  • From page
    595
  • To page
    598
  • Abstract
    An apparatus, consisting of two PbF2 crystals, each coupled to a multichannel plate photomultiplier (MCP-PMT), has been constructed in order to measure the time-of-flight (TOF) of the two 511 keV annihilation photons produced in positron emission tomography (PET). Excellent timing is achieved by detecting the prompt Cherenkov photons produced by the absorption of the 511 keV gamma photons. The present work describes the measurement and image reconstruction of two 22Na point sources. In addition, the influence of the radiator thickness and the Cherenkov light absorption cut-off of the crystal on the efficiency and the timing resolution have been studied by Monte Carlo simulation.
  • Keywords
    PET , Cherenkov radiation , Time-of-flight , Microchannel plate photomultipliers , Lead fluoride
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Serial Year
    2013
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Record number