• Title of article

    Response of a BGO detector to photon and neutron sources: simulations and measurements

  • Author/Authors

    Vincke، نويسنده , , H. and Gschwendtner، نويسنده , , E. and Fabjan، نويسنده , , C.W. and Otto، نويسنده , , T.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2002
  • Pages
    9
  • From page
    102
  • To page
    110
  • Abstract
    In this paper Monte Carlo simulations (FLUKA) and measurements of the response of a BGO detector are reported. For the measurements three low-energy photon emitters (60Co, 54Mn, 137Cs) were used to irradiate the BGO from various distances and angles. The neutron response was measured with an Am–Be neutron source. Simulations of the experimental irradiations were carried out. Our study can also be considered as a benchmark for FLUKA in terms of its reliability to predict the detector response of a BGO scintillator.
  • Keywords
    BGO scintillator , FLUKA benchmarking , Neutron and photon calibration
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Serial Year
    2002
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Record number