    The gas proportional scintillation counter/microstrip gas chamber hybrid detector with argon filling

  • Author/Authors

    Monteiro، نويسنده , , C.M.B. and Veloso، نويسنده , , J.F.C.A. and Freitas، نويسنده , , D.S.A.P. and dos Santos، نويسنده , , J.M.F and Conde، نويسنده , , C.A.N.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2002
  • Pages
    8
  • From page
    169
  • To page
    176
  • Abstract
    We present the results of a study of an argon gas proportional scintillation counter instrumented with a CsI-coated microstrip plate placed in the argon envelope. Although the measured light amplification gain and the photoelectron collection efficiency are 70% and 30–40% higher than those obtained in a similar xenon detector, the charge gain achieved in argon for the avalanches produced by the photoelectrons at the microstrip anodes is a factor of ten lower than in xenon. In both cases, these gains are limited by optical positive feedback. The energy resolution achieved for 5.9 keV X-rays was 14.8% compared to 12% in xenon.
  • Keywords
    argon , Gas scintillation counters , MSGC , X-rays , CsI photocathodes