    High-resolution and high-intensity neutron diffractometer with linear position-sensitive detector

  • Author/Authors

    Lee، نويسنده , , Chang-Hee and Moon، نويسنده , , Myung-Kook and Em، نويسنده , , V.T. and Choi، نويسنده , , Young-Nam and Oh، نويسنده , , Hwa-Suk and Nam، نويسنده , , Uk-Won، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2003
  • Pages
    9
  • From page
    353
  • To page
    361
  • Abstract
    The characteristics of a neutron diffractometer using a 3He one-dimensional position-sensitive detector (PSD) with delay line readout, a 200 (length) × 100 (height) mm2 active window and 2.5 mm spatial resolution have been studied and compared with those of the High-Resolution Powder Diffractometer (HRPD) of KAERI with 32 3He conventional tube (∅50 mm) detectors and Soller collimators (10′) before detectors. For the sample to PSD distance R=1200 mm, the PSD subtends 8° angle of 2θ and provides the resolution and the peak-to-background ratio close to that for HRPD. Time for scanning (with the same statistics) of a 0–160° interval is about 1.6 times longer and the multi-PSD system providing efficiency about 10 times higher than HRPD is proposed. Because of the small angle subtended by the PSD, the parasitic peaks from the sample environment are eliminated and operation without an oscillating radial collimator is possible. Additionally, the proposed diffractometer has an advantage for small samples.
  • Keywords
    Diffractometer , Position-sensitive neutron detector , high resolution , High intensity , Multi-PSD
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Serial Year
    2003
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Record number

    2197692