Effects of the induction-gap parameters on the signal in a double-GEM detector
Author/Authors
Guedes، نويسنده , , G.P. and Breskin، نويسنده , , A. and Chechik، نويسنده , , R. and Mِrmann، نويسنده , , D.، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2003
Pages
9
From page
305
To page
313
Abstract
A study of a dual-GEM detector coupled to a strip readout anode is described. The effects of the induction electric field and GEM-to-anode gap are presented, for an operation in atmospheric pressure Ar/CO2 (70/30) and Ar/CH4 (95/5). Visible gain and anode signal pulse-shapes, measured with 5.9 keV X-rays are presented for 1–6 mm wide induction gaps and for induction fields ranging up to 6 kV cm−1. The spatial distribution of the anode charge is provided for induction gaps of 2–12 mm. The results are useful for matching the detector parameters to the position recording circuit requirements.
Keywords
GEM , Gas electron multiplier , x-ray detection , Gaseous detector , Avalanche-induced signals