Title of article
Response of CdWO4 crystal scintillator for few MeV ions and low energy electrons
Author/Authors
Bizzeti، نويسنده , , P.G. and Carraresi، نويسنده , , L. and Danevich، نويسنده , , F.A. and Fazzini، نويسنده , , T. and Maurenzig، نويسنده , , P.R. and Taccetti، نويسنده , , F. and Taccetti، نويسنده , , N. and Tretyak، نويسنده , , V.I.، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2012
Pages
7
From page
144
To page
150
Abstract
The response of a CdWO4 crystal scintillator to protons, α particles, Li, C, O and Ti ions with energies in the range 1–10 MeV was measured. The non-proportionality of CdWO4 for low energy electrons (4–110 keV) was studied with the Compton Coincidence Technique. The energy dependence of the quenching factors for ions and the relative light yield for low energy electrons was calculated using a semi-empirical approach. Pulse-shape discrimination ability between γ quanta, protons, α particles and ions was investigated.
Keywords
Dark matter detection , CdWO4 crystal , Scintillation detectors , quenching , Non-proportionality in scintillation response , Scintillation pulse-shape