• Title of article

    Efficiency calibration and coincidence summing correction for a 4π NaI(Tl) detector array

  • Author/Authors

    Byun، نويسنده , , S.H. and Prestwich، نويسنده , , W.V. and Chin، نويسنده , , K. and McNeill، نويسنده , , F.E. and Chettle، نويسنده , , D.R.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2004
  • Pages
    12
  • From page
    674
  • To page
    685
  • Abstract
    Efficiency calibration and coincidence summing correction were performed for an array of NaI(Tl) detectors developed for in vivo neutron activation analysis. The detector array consisted of eight large NaI(Tl)s with square cross-section and the enclosing geometry of the detection produced both extremely high efficiency and a severe coincidence summing effect. The efficiency was calibrated mainly by Monte Carlo simulation with the EGS4 code. In order to obtain the coincidence summing correction factors through the simulation the coincident emission of photons was modeled. The influence of the angular correlation was explicitly investigated by adding the angle-correlated sampling routine in the simulation code. The experimental efficiencies and correction factors were measured using 137Cs, 60Co and 22Na. For both peak and total efficiencies, experiment and simulation typically agreed within 5%. The discrepancy between simulation and experiment for the coincidence summing correction factors was 15% or less. The approximate methods adopted in the measurement were analyzed in detail and estimated to be reliable for calibrating high-efficiency detectors. The calculation was finally expanded from point sources to realistic volume sources.
  • Keywords
    Monte Carlo simulation , efficiency , NaI(Tl) detector , In vivo neutron activation , Coincidence summing
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Serial Year
    2004
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Record number

    2198528