• Title of article

    Image quality evaluation of the angiography imaging system at the European synchrotron radiation facility

  • Author/Authors

    Peterzol، نويسنده , , A and Bravin، نويسنده , , A and Coan، نويسنده , , P and Elleaume، نويسنده , , H، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2003
  • Pages
    6
  • From page
    45
  • To page
    50
  • Abstract
    The evaluation of the image quality of the angiography imaging system used at the Medical Beamline of European Synchrotron Radiation Facility (ESRF) is presented. A two-line 432-pixel germanium detector is utilized in the human coronary angiography program. The performances have been measured with two monochromatic beams, with energies above and below the iodine K-edge. The large-area transfer characteristics (response curve), the spatial detail transfer performance (modulation transfer function (MTF)) and the noise properties (noise power spectrum (NPS)) have been measured and used to compute the frequency-dependent detective quantum efficiency DQE of the system.
  • Keywords
    angiography , KEDSA , DQE , Synchrotron radiation , image quality
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Serial Year
    2003
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Record number

    2200136