• Title of article

    Characterization of a pixellated CdTe detector with single-photon processing readout

  • Author/Authors

    Frِjdh، نويسنده , , C. and Graafsma، نويسنده , , H. and Nilsson، نويسنده , , H.E. and Ponchut، نويسنده , , C.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2006
  • Pages
    5
  • From page
    128
  • To page
    132
  • Abstract
    A 1 mm thick pixellated CdTe detector bonded to the MEDIPIX2 [X. Llopart, M.Campbell, R. Dinapole, D. San Segundo, E. Pemigotti, IEEE Trans. Nucl. Sci. NS-49 (5, Part 1) (2002) 2279] readout chip has been characterized using a monoenergetic microbeam at the ESRF. This is an extension of the tests previously reported in Chmeissani et al. [IEEE Trans. Nucl. Sci. NS-51(5) (2004)]. The results show that a full energy peak can be obtained when a narrow beam is focused in the center of the pixel. There is also evidence of significant charge diffusion and fluorescence. The results indicate that the charge sharing is the most important problem and will cause loss of the energy information in an imaging application. The second problem is the fluorescence which limits the number of counts in the full energy peak even for hits in the center of the pixel.
  • Keywords
    Pixel detector , CdTe , Medipix , Charge sharing , Photon counting
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Serial Year
    2006
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Record number