• Title of article

    Evaluation of the charge-sharing effects on spot intensity in XRD setup using photon-counting pixel detectors

  • Author/Authors

    Nilsson، نويسنده , , H.-E. and Mattsson، نويسنده , , C.G. and Norlin، نويسنده , , B. and Frِjdh، نويسنده , , C. and Bethke، نويسنده , , K. and De Vries، نويسنده , , R.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2006
  • Pages
    5
  • From page
    182
  • To page
    186
  • Abstract
    In this study, we examine how charge loss due to charge sharing in photon-counting pixels detectors affects the recording of spot intensity in an X-ray diffraction (XRD) setup. In the photon-counting configuration, the charge from photons that are absorbed at the boarder of a pixel will be shared between two pixels. If the threshold is high enough, these photons will not be counted whereas if it is low enough, they will be counted twice. In an XRD setup, the intensity and position of various spots should be recorded. Thus, the intensity measure will be affected by the setting of the threshold. In this study, we used a system level Monte Carlo simulator to evaluate the variations in the intensity signals for different threshold settings and spot sizes. The simulated setup included an 8keV mono-chromatic source (providing a Gaussian shaped spot) and the MEDIPIX2 photon-counting pixel detector (55 μm × 55 μm pixel size with 300μm silicon) at various detector biases. Our study shows that the charge-sharing distortion can be compensated by numerical post processing and that high resolution in both charge distribution and position can be achieved.
  • Keywords
    XRD , Medipix , Pixel detector , Photon counting , x-ray imaging
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Serial Year
    2006
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Record number

    2200977