• Title of article

    Monte Carlo efficiency transfer method for full energy peak efficiency calibration of three type HPGe detectors: A coaxial N-type, a coaxial P-type and four BEGe detectors

  • Author/Authors

    Liye، نويسنده , , Liu and Jizeng، نويسنده , , Ma and Franck، نويسنده , , Didier and de Carlan، نويسنده , , Loic and Binquan، نويسنده , , Zhang، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2006
  • Pages
    6
  • From page
    608
  • To page
    613
  • Abstract
    Monte Carlo efficiency transfer method was used to determine full energy peak efficiency of three type high-purity germanium (HPGe) detectors: a coaxial N-type, a coaxial P-type and four broad energy germanium detectors (BEGe). The comparisons between calculations and point source experiments showed that the relative deviations were mostly within ±5% for BEGe detectors in the energy range 13.9–1332 keV, within ±3% for the coaxial N-type detector in the energy range 17.5–662 keV, and within ±3% for the coaxial P-type detector in the energy range 59.5–1332 keV, respectively. An evaluation of Monte Carlo efficiency transfer method at different source-to-detector geometry configurations indicates that peak efficiency may be underestimated when point sources are placed nearer than the reference position to detector on the detector axis, and may be overestimated when point sources are placed far from the detector axis, especially for low-energy photons. Some discussions about direct Monte Carlo calculation method and Monte Carlo efficiency transfer method for full-energy peak efficiency calculation are also presented in this paper.
  • Keywords
    HPGe detector , Monte Carlo , Full-energy peak efficiency , Efficiency transfer
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Serial Year
    2006
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Record number

    2201196