Title of article
Segmented quasi-coaxial HP-Ge detectors optimized for spatial localization of the events
Author/Authors
Ripamonti، نويسنده , , Giancarlo and Pulici، نويسنده , , Paolo and Abbiati، نويسنده , , Roberto، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2006
Pages
6
From page
440
To page
445
Abstract
A methodology for the design of segmented high purity Germanium detectors is presented. Its motivation follows from the necessity of making it easier to derive fast algorithms for measuring the gamma–detector interaction position. By using our study, detector geometries can be designed, which could allow a first estimate of the interaction coordinate along the carrier drift direction by analyzing the shape of the signal of a single segment. The maximum resolution that can be achieved and the corresponding conditions for the electronics are highlighted: basic unavoidable constraints limit the resolution to around 3 mm, but this first position estimate can be used, at least in principle, as a starting point for more accurate, although computationally heavy, algorithms.
Keywords
Gamma ray detectors , Segmented detectors , HPGe , Spatial localization , Pulse shape analysis
Journal title
Nuclear Instruments and Methods in Physics Research Section A
Serial Year
2006
Journal title
Nuclear Instruments and Methods in Physics Research Section A
Record number
2202062
Link To Document