Wahl، نويسنده , , U. and Correia، نويسنده , , J.G. and Czermak، نويسنده , , A. and Jahn، نويسنده , , S.G. and Jalocha، نويسنده , , P. and Marques، نويسنده , , J.G. and Rudge، نويسنده , , A. W. Schopper، نويسنده , , F. and Soares، نويسنده , , J.C. and Vantomme، نويسنده , , A. and Weilhammer، نويسنده , , P.، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2004
Pages
12
From page
245
To page
256
Abstract
Position-sensitive detector systems, initially developed for the detection of X-rays, have been adapted for their use in electron emission channeling experiments. Each detection system consists of a 30.8×30.8 mm2 22×22-pad Si detector, either of 0.3, 0.5 or 1 mm thickness, four 128-channel preamplifier chips, a backplane trigger circuit, a sampling analog to digital converter, a digital signal processor, and a personal computer for data display and storage. The operational principle of these detection systems is described, and characteristic features such as energy and position resolution and maximum count rate, which have been determined from tests with conversion electrons and β− particles in the energy range 40–600 keV, are presented.
Keywords
Position-sensitive detectors , Si pad detectors , Electron detection , Emission channeling , Lattice location