• Title of article

    Tritium discrimination using cluster size information from a DGEM detector

  • Author/Authors

    Surette، نويسنده , , R.A. and Dubeau، نويسنده , , J.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2005
  • Pages
    8
  • From page
    433
  • To page
    440
  • Abstract
    Measurement of tritium-in-air concentrations is complicated by the presence of gamma radiation, 14C, and noble gas radionuclides. In most cases the signal from tritium, in conventional ion chamber-based monitors, is much smaller than that from equal concentrations of interfering radioactive gases. The lack of discrimination between tritium and other radioactive interferences frequently results in an overestimation of the tritium hazard. eport describes a method to effectively discriminate between tritium and other interfering airborne radioactive gases using ionization cluster size information from a double gas electron multiplier (DGEM) and a suitable readout array. The readout array consists of 121 pads (an 11×11 matrix of 4×4 mm pads on 6 mm centres) on a printed circuit board, each connected to a charge sensitive amplifier. The prototype-automated discriminating monitor typically indicated a spillover fraction, in the tritium channel, in the order of ±5% for binary mixtures of tritium and 14C, 85Kr, 133Xe or 41Ar. A custom-made dynamic cluster-fitting algorithm gives results in real-time with very good accuracy under extreme conditions.
  • Keywords
    Tritium-in-air , Cluster size , Discriminating monitor , DGEM , Readout array , tritium
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Serial Year
    2005
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Record number

    2203192