Title of article
The residual stress instrument with optimized Si(2 2 0) monochromator and position-sensitive detector at HANARO
Author/Authors
Lee، نويسنده , , Chang-Hee and Moon، نويسنده , , Myung-Kook and Em، نويسنده , , Vyacheslav T. and Choi، نويسنده , , Young-Hyun and Cheon، نويسنده , , Jong-Kyu and Nam، نويسنده , , Uk-Won and Kong، نويسنده , , Kyung-Nam، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2005
Pages
10
From page
480
To page
489
Abstract
An upgraded residual stress instrument at the HANARO reactor of the KAERI is described. A horizontally focusing bent perfect crystal Si(2 2 0) monochromator (instead of a mosaic vertical focusing Ge monochromator) is installed in a drum with a tunable (2θM=0–60°) take-off angle/wavelength. A specially designed position-sensitive detector (60% efficiency for λ = 1.8 A ) with 200 mm (instead of 100 mm) high-active area is used. There are no Soller type collimators in the instrument. The minimum possible monochromator to sample distance, L MS = 2 m , and sample to detector distance, L SD = 1.2 m , were found to be optimal. The new PSD and bent Si(2 2 0) monochromator combined with the possibility of selecting an appropriate wavelength resulted in about a ten-fold gain in data collection rate.
timal reflections of austenitic and ferritic steels, aluminum and nickel for stress measurements with a Si(2 2 0) monochromator were chosen experimentally. The ability of the instrument to make strain measurements deep inside the austenitic and ferritic steels has been tested. For the chosen reflections and wavelengths, no shift of peak position (apparent strain) was observed up to 56 mm length of path.
Keywords
Residual Stress Measurement , Bent perfect silicon crystal , Monochromator , Position-sensitive detector , VAMAS , Neutron diffraction
Journal title
Nuclear Instruments and Methods in Physics Research Section A
Serial Year
2005
Journal title
Nuclear Instruments and Methods in Physics Research Section A
Record number
2203651
Link To Document