CdZnTe detectors for small field of view positron emission tomographic imaging
Author/Authors
Drezet، نويسنده , , Arnaud and Monnet، نويسنده , , Olivier and Mathy، نويسنده , , Françoise and Montemont، نويسنده , , Guillaume and Verger، نويسنده , , Loïck، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2007
Pages
6
From page
465
To page
470
Abstract
We report the timing performance between two 16×20×0.9 mm3 cadmium zinc telluride (CZT) detectors equipped with a specific orthogonal electrode geometry (16 anodes and 5 cathodes with 1 and 4 mm pitch, respectively) and a dedicated electronic setup for Positron Emission Tomography (PET) application. The measured coincidence times reach 2.6 ns full-width at half-maximum (FWHM) for a 500 V bias voltage and 300 keV energy threshold. Subsequently, a simulation study was carried out to assess the spatial and efficiency performances of these detectors, which allow the depth of interaction (DOI) measurement and so limit the parallax error. Preliminary results show that the proposed design could reach a better homogeneity of the spatial resolution across the field of view than what is achieved with the standard PET scintillation.
Keywords
Timing resolution , Positron emission tomography , CdZnTe , Depth of interaction measurement