• Title of article

    Automation of radiation dosimetry using PTW dosemeter and LabVIEW™

  • Author/Authors

    Weiss، نويسنده , , C. and Al-Frouh، نويسنده , , K. and Anjak، نويسنده , , O.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2011
  • Pages
    4
  • From page
    649
  • To page
    652
  • Abstract
    Automation of UNIDOS “Dosemeter” using personal computer (PC) is discussed in this paper. er to save time and eliminate human operation errors during the radiation dosimetry, suitable software, using LabVIEW™ graphical programming language, was written to automate and facilitate the processes of measurements, analysis and data storage. ftware calculates the calibration factor of the ionization chamber in terms of air kerma or absorbed dose to water according to IAEA dosimetry protocols. It also has the ability to print a calibration certificate. tained results using this software are found to be more reliable and flexible than those obtained by manual methods previously employed. LabVIEW™ as a development tool is extremely convenient to make things easier when software modifications and improvements are needed.
  • Keywords
    Radiation dosimetry , LabVIEW™ , Ion chamber calibration
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Serial Year
    2011
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Record number