• Title of article

    Modification of a medical PET scanner for PEPT studies

  • Author/Authors

    Sadrmomtaz، نويسنده , , Alireza and Parker، نويسنده , , D.J. and Byars، نويسنده , , L.G.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2007
  • Pages
    4
  • From page
    91
  • To page
    94
  • Abstract
    Over the last 20 years, positron emission tomography (PET) has developed as the most powerful functional imaging modality in medicine. Over the same period the University of Birmingham Positron Imaging Centre has applied PET to study engineering processes and developed the alternative technique of positron emission particle tracking (PEPT) in which a single radioactively labelled tracer particle is tracked by detecting simultaneously the pairs of back-to-back photons arising from positron/electron annihilation. Originally PEPT was performed using a pair of multiwire detectors, and more recently using a pair of digital gamma camera heads. In 2002 the Positron Imaging Centre acquired a medical PET scanner, an ECAT 931/08, previously used at Hammersmith Hospital. This scanner has been rebuilt in a flexible geometry for use in PEPT studies. This paper presents initial results from this system. Fast moving tracer particles can be rapidly and accurately located.
  • Keywords
    PEPT , BGO detectors , Particle tracking
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Serial Year
    2007
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Record number