• Title of article

    Wide-window gas target system for high resolution experiment with magnetic spectrometer

  • Author/Authors

    Matsubara، نويسنده , , H. and Tamii، نويسنده , , A. and Shimizu، نويسنده , , Y. and Suda، نويسنده , , K. and Tameshige، نويسنده , , Y. and Zenihiro، نويسنده , , J.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2012
  • Pages
    8
  • From page
    122
  • To page
    129
  • Abstract
    A gas target system has been developed for nuclear scattering experiments with a highly dispersive spectrometer at small scattering angles including 0°. The system is equipped with a gas cell that has a wide beam window of 44 mm in width and 14 mm in height. The size is sufficient to pass a dispersed beam, which is essential for high energy-resolution measurement. A gas handling system allows target gas that is often limited in quantity, to be collected for reuse. Typical areal density is 1 mg/cm2 for 20Ne or 36Ar gas. The target can be operated either at room temperature or at the liquid nitrogen temperature. The gas target has been employed in a proton inelastic-scattering experiment at 295 MeV with the Grand Raiden spectrometer at the Research Center for Nuclear Physics. A good energy resolution of 20 keV (FWHM) has been achieved at scattering angles of 0–4.5°.
  • Keywords
    Zero degree measurement , Dispersion matching , Target cell , Gas target , Reusable gas , High energy-resolution
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Serial Year
    2012
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Record number