• Title of article

    Monte Carlo optimization of sample dimensions of an 241Am–Be source-based PGNAA setup for water rejects analysis

  • Author/Authors

    Idiri، نويسنده , , Z. and Mazrou، نويسنده , , H. and Beddek، نويسنده , , S. and Amokrane، نويسنده , , A. and Azbouche، نويسنده , , A.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2007
  • Pages
    10
  • From page
    279
  • To page
    288
  • Abstract
    The present paper describes the optimization of sample dimensions of a 241Am–Be neutron source-based Prompt gamma neutron activation analysis (PGNAA) setup devoted for in situ environmental water rejects analysis. The optimal dimensions have been achieved following extensive Monte Carlo neutron flux calculations using MCNP5 computer code. A validation process has been performed for the proposed preliminary setup with measurements of thermal neutron flux by activation technique of indium foils, bare and with cadmium covered sheet. Sensitive calculations were subsequently performed to simulate real conditions of in situ analysis by determining thermal neutron flux perturbations in samples according to chlorine and organic matter concentrations changes. The desired optimal sample dimensions were finally achieved once established constraints regarding neutron damage to semi-conductor gamma detector, pulse pile-up, dead time and radiation hazards were fully met.
  • Keywords
    PGNAA technique , Optimal sample dimensions , Monte Carlo calculations , MCNP5 , Semi-conductor gamma detector , activation foil technique
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Serial Year
    2007
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Record number

    2206669