• Title of article

    Aspects of GEANT4 Monte-Carlo calculations of the BC501A neutron detector

  • Author/Authors

    Patronis، نويسنده , , N. and Kokkoris، نويسنده , , M. and Giantsoudi، نويسنده , , D. and Perdikakis، نويسنده , , G. and Papadopoulos، نويسنده , , C.T. and Vlastou، نويسنده , , R.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2007
  • Pages
    5
  • From page
    351
  • To page
    355
  • Abstract
    The response function of a commercial 2 in.×2 in. BC501A neutron detector has been studied through the general-purpose Monte-Carlo (MC) simulation toolkit GEANT4. For neutron energies below 8 MeV, where the dominant detection mechanism originates from the knockout protons, the results of MC calculations reproduce remarkably well the experimental pulse-height spectra. This is not the case, however, at higher neutron energies, where the contribution of heavier charged particles to the detection mechanism becomes significant. The calculated detector efficiency is also compared at three indicative energies to the experimental one via the neutron activation technique.
  • Keywords
    neutron spectroscopy , Scintillation detectors , Monte-Carlo simulations , GEANT4
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Serial Year
    2007
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Record number

    2206687