• Title of article

    Sensitivity assessment of wide Axial Field of View PET systems via Monte Carlo simulations of NEMA-like measurements

  • Author/Authors

    Couceiro، نويسنده , , M. and Ferreira، نويسنده , , N.C. and Fonte، نويسنده , , P.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2007
  • Pages
    4
  • From page
    485
  • To page
    488
  • Abstract
    The sensitivity characteristics of Positron Emission Tomography (PET) systems with wide Axial Field of View (AFOV) was studied by MonteCarlo simulations complemented by an approximate analytical model, aiming at full-body human PET systems with AFOV in the order of 200 cm. Simulations were based on the GEANT4 package and followed closely the NEMA NU-2 1994 norm. The sensitivity, dominated by the solid angle, grows strongly with the AFOV and with the axial acceptance angle, while the scatter fraction is almost independent from the geometry.
  • Keywords
    PET , Sensitivity , Simulation
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Serial Year
    2007
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Record number

    2207542