• Title of article

    A simple X-ray position detection system based on a MHSP

  • Author/Authors

    Natal da Luz، نويسنده , , H. and Veloso، نويسنده , , J.F.C.A. and dos Santos، نويسنده , , J.M.F. and Mir، نويسنده , , J.A.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2007
  • Pages
    4
  • From page
    1083
  • To page
    1086
  • Abstract
    A simple system for position detection based on a Micro-Hole and Strip Plate (MHSP) using a resistive charge division method was implemented. A thin resistive layer of about 100 Ω between each strip (top surface) and between each anode strip (bottom surface) allows us to obtain the actual position in both orthogonal directions. The readout electronics uses only two preamplifiers, a sum and a divider circuit to calculate the interaction point, in each dimension. For a detector filled with 1 atm of pure xenon, position resolutions of less than 300 μm (FWHM) were achieved using 5.9 keV X-rays, for a signal-to-noise ratio (SNR) of 30. For these conditions, a detector gain of about 104 and an energy resolution of about 15% were obtained.
  • Keywords
    Micro-hole and strip plate , 2D imaging , Resistive charge division , X-ray detectors , Analogue divider , Gas electron multiplier
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Serial Year
    2007
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Record number

    2207847